<ErrorNumber>15309</ErrorNumber>
<ErrorFamily>Autoloader</ErrorFamily>
<ErrorMessage>
ALE: Rotator move timed out moving to rotated|unrotated position
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Trendsetter_400/800_Auto</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<OperatorAction>
<DescriptionText><text>
Select the RESUME button on the workstation dialog box. If this fails to correct
the problem, check that nothing is obstructing the unload table traveller from
rotating. If everything looks normal, then Creo Service should be called.
</text></DescriptionText>
</OperatorAction>
<ServiceInformation>
<DescriptionText><text>
The unload table traveller rotate sensors failed to trip within the timeout
specified by the aldly ucr parameter. The sensor status may be checked by using
the TABLE command. The sensor status should read TRUE if the sensors are
activated.
</text></DescriptionText>
<DescriptionText><text>
It may be useful to set the 'verbose alcmd' and 'verbose alconv' parameters to 1
in order to obtain further diagnostic information.
</text></DescriptionText>
<DescriptionText><text>
Typing "ALE" will display the system air pressure.
</text></DescriptionText>
</ServiceInformation>
<Cause>
The rotator may be obstructed by something.
</Cause>
<Cause>
The rotator magnetic sensors may not be adjusted correctly.
</Cause>
<Cause>
The magnetic sensors may have been damaged.
</Cause>
<Cause>
The aldly ucr timeout value may be set too short.
</Cause>
<Cause>
The wiring to the air valve solenoid may be disconnected or damaged.
</Cause>
<Cause>
The wiring to the magnetic sensors may be disconnected or damaged.
</Cause>
<Cause>
The pneumatic speed controls may be mis-adjusted.
</Cause>
<Cause>
The pneumatic system may be damaged.
</Cause>
<Cause>
The air supply may be low or turned off.
</Cause>
<Cause>
If the unload table traveller rotates properly but the sensors are not
working, then the sensors may be temporarily masked off by setting al urs to 0.
Masking off these sensors will get rid of the error message but will cause the
machine to run significantly slower than it used to.
</Cause>
<Cause>
WARNING:  Masking off sensors disables proper error checking!
</Cause>
<SeeAlso>
<UnnumberedList>
<Link><LinkAddress>../commands/table.xml</LinkAddress><LinkTitle>Command: table</LinkTitle></Link>
<Link><LinkAddress>../commands/resume.xml</LinkAddress><LinkTitle>Command: resume</LinkTitle></Link>
<Link><LinkAddress>../commands/ale.xml</LinkAddress><LinkTitle>Command: ale</LinkTitle></Link>
<Link><LinkAddress>../nvs/al_cal_urs.xml</LinkAddress><LinkTitle>NVS: al urs</LinkTitle></Link>
<Link><LinkAddress>../nvs/aldly_cal_ucr.xml</LinkAddress><LinkTitle>NVS: aldly ucr</LinkTitle></Link>
<Link><LinkAddress>../nvs/verbose_cfg_alcmd.xml</LinkAddress><LinkTitle>NVS: verbose alcmd</LinkTitle></Link>
<Link><LinkAddress>../nvs/verbose_cfg_alconv.xml</LinkAddress><LinkTitle>NVS: verbose alconv</LinkTitle></Link>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:00 2008 PST
</Updated>
<Links>
</Links>
</Error>